Thermocouples – measuring insert (data sheet 8666)

Intrinsically safe


The measuring inserts TTeMiXiAo and TTeMiXiAoT according to DIN 43 735 are solely intended for installation into protective fittings for electrical thermometers with connection heads form B according to DIN EN 50 446 and have an EU Type Examination Certificate for the ignition protection type “intrinsic safety”. Both models meet the requirements of the directive 2014/34/EU for the application in potentially explosive atmospheres due to gases and dusts.
For the operation in connection heads with flameproof enclosure (types XD-AD and XD-SD), the measuring inserts are optionally available with a sleeve, which forms, in combination with a socket, a flameproof gap in the connection head.
For both models, we offer various thermocouples according to DIN EN 60 584. In addition, model TTeMiXiAoT is available with several fitted transmitters with analogue or digital output.
The measuring inserts shall only be used in a suitable housing.
 
8666 Measuring insert TTeMiXiAo replacement part for the installation into connection heads ARMANO

8666 Measuring insert replacement part for the installation into connection heads ARMANO
model TTeMiXiAo / TTeMiXiAoT
operating temperature range type K (NiCr-Ni):         −40 / +1175 °C (−40 / +2147 °F)1)
type N (NiCrSi-NiSi): −40 / +1175 °C (−40 / +2147 °F)1)
type J (Fe-CuNi):        −40 /   +750 °C (−40 / +1382 °F)
type S (Pt10Rh-Pt):        0  / +1175 °C (+32 / +2147 °F)1)

1)
application range limited due to the sheath material Inconel 600
degree of protection (DIN EN 60 529) IP00
measuring inserts are intended for the installation into protective fittings for electrical thermometers, which are equipped with an appropriate degree of protection for safe operation
approvals II 1G Ex ia IIC T6…T1 Ga2)
II 2G Ex ia IIC T6…T1 Gb2)
II 1D Ex ia IIIC T80 °C…T440 °C Ga2)
II 2D Ex ia IIIC T80 °C…T440 °C Gb2)

2) only with suitable protective fitting
application in explosion hazardous areas zones 0, 1, 2:        applicable with protective fitting with degree of protection of at least IP54
zones 20, 21, 22: applicable with mounting part with degree of protection of at least IP6X
output signal model TTeMiXiAo:   thermoelectric voltage according to DIN EN 60 584
model TTeMiXiAoT: 4…20 mA, HART® or PROFIBUS® PA / FOUNDATION™ Fieldbus