Resistance thermometer with process display (data sheet 8590)

With flameproof enclosure, intrinsically safe


The resistance thermometer model TPtPAXd is designed for the installation into solid drilled or fabricated thermometer thermowells (e.g. according to DIN 43 772). Without thermowell, this version must not be applied in pressurised media.
TPtPAXd is equipped with a rugged die-cast aluminum case with an integrated backlit LC-graphic display, which communicates digitally with the fitted transmitter type 5337-D via HART interface.
Both digital display and transmitter can be parameterised via optical keyboard through the window of the instrument without having to remove the screw-on lid.
TPtPAXd has an EU Type Examination Certificate for the ignition protection types “flameproof enclosure” and “intrinsic safety” and meets the requirements of the directive 2014/34/EU for the application in potentially explosive atmospheres due to gases and dusts.
 
8590 Resistance thermometer TPtPAXd with process display ARMANO

model TPtPAXd
operating temperature range −200 / +600 °C (−328 / +1112 °F)
degree of protection (DIN EN 60 529) IP66 – 68
when mounted in a thermowell, depending on the screwed cable gland used
approvals II 1G Ex ia db IIC T6…T1 Ga
II 1/2G Ex db IIC T6…T1 Ga/Gb1)
II 1/2G Ex ia/db IIC T6…T1 Ga/Gb1)
II 2G Ex db IIC T6…T1 Gb
II 1D Ex ia tb IIIC T80 °C…T440 °C Da
II 1/2D Ex ia/tb IIIC T80 °C…T440 °C Da/Db1)
II 2D Ex tb IIIC T80 °C…T440 °C Db

1) only with thermowell suitable for zone separation – see operating instructions B71
application in explosion hazardous areas zones 1, 2, 21, 22
zones 0, 20: applicable with measuring insert with ignition protection type Ex ia
output signal 4…20 mA, HART®